COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered dilu...
FDA Device Recall #Z-1616-2015 — Class II — April 13, 2015
Recall Summary
| Recall Number | Z-1616-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 13, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Beckman Coulter Inc. |
| Location | Brea, CA |
| Product Type | Devices |
| Quantity | 829,888 units total (659,584 units in US) |
Product Description
COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.
Reason for Recall
Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.
Distribution Pattern
Worldwide Distribution -- Argentina; Brazil; Brunei Darussalam; Canada; Chile; Colombia; Hong Kong; India; Lebanon; Macao; Mexico; Panama; Philippines; Qatar; Singapore; Taiwan; Thailand; Trinidad and Tobago; United States, including Puerto Rico; and Uruguay.
Lot / Code Information
Lot No. 3099980 - 3512190, expiration dates: 17 April 2015 - 26 August 2016.
Other Recalls from Beckman Coulter Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0541-2026 | Class II | DxFLEX Flow Cytometer Catalog Number C78500 UD... | Sep 22, 2025 |
| Z-0441-2026 | Class II | SYNCHRON Systems Direct LDL Cholesterol (LDLD) ... | Sep 17, 2025 |
| Z-0071-2026 | Class II | SYNCHRON Systems Phosphorus (PHOSm) Reagent RE... | Sep 3, 2025 |
| Z-2573-2025 | Class II | UniCel DxH 690T COULTER Cellular Analysis Syste... | Aug 13, 2025 |
| Z-2574-2025 | Class II | UniCel DxH 800 COULTER Cellular Analysis System... | Aug 13, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.