PENTRA C200, Part Number: A11A01629. The Pentra C200 system is a fully automated chemistry ana...
FDA Device Recall #Z-0633-2013 — Class II — October 10, 2012
Recall Summary
| Recall Number | Z-0633-2013 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 10, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Horiba Instruments, Inc dba Horiba Medical |
| Location | Irvine, CA |
| Product Type | Devices |
| Quantity | 384 units |
Product Description
PENTRA C200, Part Number: A11A01629. The Pentra C200 system is a fully automated chemistry analyzer using colorimetry, turbidmetry and potentiometry technologies. It is mostly dedicated to in vitro diagnostic analyses based on homogeneous samples such as serum, plasma and uring.
Reason for Recall
Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.
Distribution Pattern
Nationwide Distribution including CA, OK, NY, FL, AZ, NY, VA, and TX.
Lot / Code Information
Lot number: 00456-4108
Other Recalls from Horiba Instruments, Inc dba Horiba Me...
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2286-2018 | Class II | ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, RE... | Apr 26, 2018 |
| Z-0493-2018 | Class II | 10 ML REAGENT CUP, B1034626, SAP 1221034626 for... | Nov 21, 2017 |
| Z-0494-2018 | Class II | 15 ML REAGENT CUP, B1037307, SAP 1221037307 for... | Nov 21, 2017 |
| Z-0902-2015 | Class II | ABX PENTRA Reagent Container, Model No. B103730... | Nov 13, 2014 |
| Z-2692-2014 | Class II | ABX PENTRA Magnesium RTU Reagent, Part No. A11A... | Aug 11, 2014 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.