Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for...
FDA Device Recall #Z-2185-2012 — Class I — June 29, 2012
Recall Summary
| Recall Number | Z-2185-2012 |
| Classification | Class I — Serious risk |
| Date Initiated | June 29, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | CareFusion 303, Inc. |
| Location | San Diego, CA |
| Product Type | Devices |
| Quantity | 102,248 units |
Product Description
Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for facilities that utilize infusion for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), intra-arterial (IA),subcutaneous, epidural, enteral, or irrigation of fluid spaces. The Pump module is indicated for use on adults, pediatrics, and neonates.
Reason for Recall
The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with Alaris Pump Module (model 8100) manufactured between August 2010 and July 2011.
Distribution Pattern
Worldwide Distribution - USA (nationwide) and the countries of: South Africa and Canada. Additional Countries: Netherlands.
Lot / Code Information
All serial numbers associated with the Alaris Pump Module Model 8100: 8100 DEMO PUMP MODULE; 8100 Global Pump Module; 8100 IOC Pump Module V8.5.29.0; 8100 LVP Training Unit; 8100 Pump Module V8.5.29.0; 8100 Pump Module V8.5.29.0 (V9); Medley Patient Care System LVP; Model 8100 Trade-In Material; RCND 8100 Pump Module V8.5.29.0
Other Recalls from CareFusion 303, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1645-2026 | Class II | BD Alaris" System with Guardrails" Suite MX wit... | Feb 11, 2026 |
| Z-1028-2026 | Class II | BD Pyxis ES Enterprise Server Catalog Numbers 1... | Dec 3, 2025 |
| Z-0962-2026 | Class II | BD Pyxis MedStation ES, REF: 323, BD Pyxis M... | Nov 20, 2025 |
| Z-1217-2026 | Class II | BD Alaris" Pump Module model 8100 UDI-DI code:... | Nov 6, 2025 |
| Z-0430-2026 | Class I | BD Alaris infusion Pump Module 8100, REF: 8100;... | Oct 17, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.