Comfort brand Ear-Loop Face Mask, Item #: 3100, Model Numbers: # 3100B, # 3100P, # 3100G, # 3100Y...

FDA Device Recall #Z-1451-2013 — Class II — May 15, 2013

Recall Summary

Recall Number Z-1451-2013
Classification Class II — Moderate risk
Date Initiated May 15, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Maytex Corp
Location Hayward, CA
Product Type Devices
Quantity 43,915 Cases

Product Description

Comfort brand Ear-Loop Face Mask, Item #: 3100, Model Numbers: # 3100B, # 3100P, # 3100G, # 3100Y, # 3100W; Product is manufactured and distributed by Maytex Corporation, CA Ear-loop face mask for basic medical/dental procedures.

Reason for Recall

Recent Lab testing conducted in March 2013 indicates that the product does not meet the standard of "Differential Pressure of only 1.6 Delta P (mm H2O/cm2)" as indicated on the box/package of the product; and also product manufactured between 2008 - 2012 cannot be guaranteed as Fluid Resistant.

Distribution Pattern

Worldwide distribution: USA (nationwide) including states of: AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, MA, MD, ME, MI, MN, MO, MT, NE, NH, NJ, NM, NY, OK, OR, PA, PR, TN, TX, UT, WA, and WI; and country of: Mexico.

Lot / Code Information

All cases of item #: 3100 face mask with PO #. PO #, Date of receipt: 809-01, 2/5/08; 837-02, 7/29/08; 852-03, 10/10/08; 901-01, 2/3/09; 921-02, 6/8/09; 923-03, 7/8/09; 926-04, 8/26/09; 931-05, 11/10/09; 942-06, 12/16/09; 001-01, 2/17/10; 003-03, 4/12/10; 002-02 , 3/24/10; 003-03A, 7/23/10; 042-04, 11/29/10; 105-01, 1/31/11; 117-02, 4/29/11; 127/03, 7/8/11; 147-04, 12/16/11; 202-01, 3/8/2012; 224-02, 5/24/2012; 235-03, 9/12/2012; 253-04 , 12//192012

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.