GE Healthcare Millennium MC

FDA Device Recall #Z-0371-2019 — Class I — September 14, 2018

Recall Summary

Recall Number Z-0371-2019
Classification Class I — Serious risk
Date Initiated September 14, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm GE Healthcare, LLC
Location Waukesha, WI
Product Type Devices
Quantity 18

Product Description

GE Healthcare Millennium MC

Reason for Recall

A detector can detach and fall.

Distribution Pattern

United States: AK,AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,Guam,MA,MD,ME,MI,MO,MS,MT,NC,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR, RI,SC,SD,TN,TX,UT,VA,VT,WA,WI AND WV International: Albania, Argentina, Australia, Azerbaijan, Belgium, Bosnia ,Herzegovina, Brazil, Canada, Chile, China, Costa Rica, Colombia, Cyprus, France, Germany, Greece, Guatemala, Honduras, Iceland, India, Iraq, Italy, Jamaica, Japan, Kenya, Korea, Macedonia, Mexico, Namibia, Panama, Peru, Philippines, Portugal, Russia, South Africa, Spain Switzerland, Taiwan, United Kingdom, Turkey, Venezuela and Vietnam

Lot / Code Information

System ID #: 870735MG, 928333WMM02, 970635MYO, 203576DNT, 407896FMG2, 954634MG, 316267KCN, 601883NMC2, 910483XEL1, 910296MMG, 908277XL2, 908277MG2, 908218MYO, 918496EN1, 972759CROWNMG, 940566MG, 972434MC,and 972473MYO.

Other Recalls from GE Healthcare, LLC

Recall # Classification Product Date
Z-1566-2022 Class II (1) GE Centricity Universal Viewer 6.0, Model 2... Jul 15, 2022
Z-1430-2022 Class II SIGNA Premier magnetic resonance scanner, model... Jun 24, 2022
Z-1466-2022 Class II Centricity PACS Software Version 7.0 SP0.0.4.7 Jun 14, 2022
Z-1333-2022 Class II GE Centricity Universal Viewer Zero Footprint. ... Jun 8, 2022
Z-1511-2022 Class II Service manuals for the CT, PET, or NM table as... Jun 2, 2022

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.