Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
FDA Device Recall #Z-1844-2024 — Class II — May 3, 2024
Recall Summary
| Recall Number | Z-1844-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 3, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Philips North America Llc |
| Location | Cambridge, MA |
| Product Type | Devices |
Product Description
Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
Reason for Recall
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Lot / Code Information
UDI-DI: N/A Serial Numbers: 42246 85713 42029 71774 85668 85680 85685 85694 85705 42023 42024 42375 42217 42402 71044 85698 85700 85146 85677 85699 71595 71815 85649 42320 85684 71836 85679 85682 85697 85701 85708 85714
Other Recalls from Philips North America Llc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1316-2026 | Class II | IQon Spectral CT; Product Code (REF): 728332; | Jan 7, 2026 |
| Z-1318-2026 | Class II | Spectral CT 7500 on Rails; Product Code (REF):... | Jan 7, 2026 |
| Z-1317-2026 | Class II | Spectral CT; Product Code (REF): 728333; | Jan 7, 2026 |
| Z-1315-2026 | Class II | Brilliance iCT; Product Code (REF): 728306; | Jan 7, 2026 |
| Z-1033-2026 | Class II | Philips Smart-hopping 2.0 AP 1.4 GHz. Patient M... | Dec 12, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.