Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches o...

FDA Recall #Z-0267-2013 — Class II — October 13, 2008

Recall #Z-0267-2013 Date: October 13, 2008 Classification: Class II Status: Terminated

Product Description

Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.

Reason for Recall

Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.

Recalling Firm

Medacta Usa Inc — Camarillo, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

12 units

Distribution

Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.

Code Information

Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243)

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated