Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light Syste...
FDA Device Recall #Z-0563-2016 — Class II — November 2, 2015
Recall Summary
| Recall Number | Z-0563-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 2, 2015 |
| Status | Completed |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Trumpf Medical Systems, Inc. |
| Location | Charleston, SC |
| Product Type | Devices |
| Quantity | 20,713 |
Product Description
Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5
Reason for Recall
The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.
Distribution Pattern
USA (nationwide) Distribution to the states of : AL, AK, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KY, ME, MA, MN, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WY., and Internationally to Canada.
Lot / Code Information
4028110 - iLED Single; 4028210 - iLED duo; 4028310 - iLED trio/quad; 4038110 - TruLight single; 4038210 - TruLight duo; 4038310 - TruLight trio/quad; 1565068 - iLED single (TUSM); 1565160 - iLED duo (TUSM); 1565161 - iLED trio/quad (TUSM); 1574759 - TruLight single (TUSM); 1574850 - TruLight duo (TUSM); 1574851 - TruLight trio/quad (TUSM)
Other Recalls from Trumpf Medical Systems, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0824-2018 | Class II | Camera System, Television, Surgical without aud... | Dec 15, 2017 |
| Z-0391-2017 | Class II | OR Table, Top U26 H V, Model # 1909793; OR Tabl... | Oct 25, 2016 |
| Z-0043-2017 | Class II | Operating Table TruSystem 7000 U, Material Numb... | Jul 10, 2016 |
| Z-0040-2017 | Class II | Mobile Column TruSystem 7500, Material Number 1... | Jul 10, 2016 |
| Z-0042-2017 | Class II | Floor Mounting Column TS 7500 U, Material Numbe... | Jul 10, 2016 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.