Dimension Vista(R) IRON Flex(R) reagent cartridge (K3085) The IRON method is an in vitro diagn...
FDA Device Recall #Z-0452-2013 — Class II — October 8, 2012
Recall Summary
| Recall Number | Z-0452-2013 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 8, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Healthcare Diagnostics, Inc. |
| Location | Newark, DE |
| Product Type | Devices |
| Quantity | 19,310 |
Product Description
Dimension Vista(R) IRON Flex(R) reagent cartridge (K3085) The IRON method is an in vitro diagnostic test for the quantitative determination of iron in human serum and plasma on the Dimension Vista(R) System. Measurements of iron are used in the diagnosis and treatment of diseases such as iron deficiency anemia and other disorders of iron metabolism.
Reason for Recall
Siemens Healthcare Diagnostics has confirmed the potential for falsely elevated results with IRON Flex(R) reagent cartridges when tests are processed from a well set contaminated with iron from environmental sources. All lots of IRON Flex(R) reagent cartridges have the potential for this issue. If a well set is contaminated with trace amounts of iron, the entire well set is impacted, and the mag
Distribution Pattern
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, Washington DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and Internationally to Canada, AU, NZ, JP, DE, SA, and MY
Lot / Code Information
Lot numbers 11286BA, 11319BB, 12004BB, 12044BC, 12073BB, 12073BC, 12101BC, 12101BD, 12130BB, 12130BC and 12188BD.
Other Recalls from Siemens Healthcare Diagnostics, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1484-2026 | Class II | Atellica CH Urine Albumin (UAlb). Material Numb... | Jan 20, 2026 |
| Z-0491-2026 | Class II | IMMULITE 2000 Oak Mix Specific Allergen; Sieme... | Oct 6, 2025 |
| Z-0492-2026 | Class II | 3gAllergy Specific IgE Universal Kit; Kit Comp... | Oct 6, 2025 |
| Z-0493-2026 | Class II | Food Panel 3 Specific Allergen; Siemens Materi... | Oct 6, 2025 |
| Z-0366-2026 | Class II | Assay: IMMULITE 2000 Intact PTH; Test Code: iP... | Sep 22, 2025 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.