3.2MM X 560MM COCR THD TIP; 14-441053 Intended to be used as a guide pin for insertion of imp...

FDA Device Recall #Z-2242-2018 — Class II — May 16, 2018

Recall Summary

Recall Number Z-2242-2018
Classification Class II — Moderate risk
Date Initiated May 16, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Zimmer Biomet, Inc.
Location Warsaw, IN
Product Type Devices
Quantity 2396

Product Description

3.2MM X 560MM COCR THD TIP; 14-441053 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.

Reason for Recall

The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.

Distribution Pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.

Lot / Code Information

631960 668040 668020 668030 668050 668060 265460 265450 322270 354190 370190 918470 034780 101840 216280 135720 327770 363260 389080 389070 399500 436760 436770 471160 471170 005620 244600 403100 791030 791110 839790 898930 898940 898960 974090 974100 974060 974070 974080 038580 142780 142790 492260 522530 205310 767140 492270 341690 620450 118270 418780 107090 373830 600950 609070 044210 044220 612900 683640 683630 507830 675290 284100 132110 962210

Other Recalls from Zimmer Biomet, Inc.

Recall # Classification Product Date
Z-0696-2021 Class II Spinal Rod Cutter, Item number 00392500200. The... Nov 20, 2020
Z-0570-2021 Class II Polaris 5.5 Spinal System and Polaris 5.5/Cyphe... Nov 17, 2020
Z-0372-2021 Class II ZCA All Poly Acetabular Cup Longevity Crosslink... Sep 24, 2020
Z-0345-2021 Class II ZCA All Poly Acetabular Cup Longevity Crosslink... Sep 24, 2020
Z-0347-2021 Class II ZCA All Poly Acetabular Cup Longevity Crosslink... Sep 24, 2020

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.