LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

FDA Device Recall #Z-2445-2024 — Class II — June 27, 2024

Recall Summary

Recall Number Z-2445-2024
Classification Class II — Moderate risk
Date Initiated June 27, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm LINK BIO CORP
Location Dover, NJ
Product Type Devices
Quantity 8 units

Product Description

LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

Reason for Recall

The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.

Distribution Pattern

US Nationwide distribution in the states of FL, GA, KS, OH, WI.

Lot / Code Information

Item Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066.

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.