Product Description
CLARAVUE Pre-wired Electrode - IP SET 12 RT ADULT-US Model: 32028779 50602-US
Reason for Recall
Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.
Distribution Pattern
US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.
Lot / Code Information
All lots UDI: 03700506307326
Other Recalls from Graphic Controls Acquisition Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0936-2022 | Class II | CLARAVUE Pre-wired Electrode - IP SET M3 RT ADU... | Mar 15, 2022 |
| Z-0940-2022 | Class II | CLARAVUE Pre-wired Electrode - IP SET 5V RT AD... | Mar 15, 2022 |
| Z-0937-2022 | Class II | CLARAVUE Pre-wired Electrode - IP SET M5 RT ADU... | Mar 15, 2022 |
| Z-0939-2022 | Class II | CLARAVUE Pre-wired Electrode - IP SET 6V RT AD... | Mar 15, 2022 |
| Z-0935-2022 | Class II | CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIAT... | Mar 15, 2022 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.