Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 16 System Number:...
FDA Device Recall #Z-0280-2022 — Class II — May 1, 2021
Recall Summary
| Recall Number | Z-0280-2022 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 1, 2021 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Philips North America Llc |
| Location | Cambridge, MA |
| Product Type | Devices |
| Quantity | 3 units |
Product Description
Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 16 System Number: 882470 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.
Reason for Recall
Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part
Distribution Pattern
US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to countries of: Australia, Germany, Netherlands and Saudi Arabia.
Lot / Code Information
Serial Number: 7084 7145 7584
Other Recalls from Philips North America Llc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1316-2026 | Class II | IQon Spectral CT; Product Code (REF): 728332; | Jan 7, 2026 |
| Z-1318-2026 | Class II | Spectral CT 7500 on Rails; Product Code (REF):... | Jan 7, 2026 |
| Z-1317-2026 | Class II | Spectral CT; Product Code (REF): 728333; | Jan 7, 2026 |
| Z-1315-2026 | Class II | Brilliance iCT; Product Code (REF): 728306; | Jan 7, 2026 |
| Z-1033-2026 | Class II | Philips Smart-hopping 2.0 AP 1.4 GHz. Patient M... | Dec 12, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.