PKG, 5MM INSERT, ENDO METZENBAUM SCISSORS, CURVED 45CM, P/N 0250080575 Laparoscopic Manual Instr...
FDA Device Recall #Z-0703-2015 — Class II — November 17, 2014
Recall Summary
| Recall Number | Z-0703-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 17, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Stryker Endoscopy |
| Location | San Jose, CA |
| Product Type | Devices |
| Quantity | 180,573 units total |
Product Description
PKG, 5MM INSERT, ENDO METZENBAUM SCISSORS, CURVED 45CM, P/N 0250080575 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason for Recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lot / Code Information
247425 345126 443972 542154 643084 744694 745080 745631 746499 747596 845622 940158 941057 942853 0943888E 0943889E 0945145E 0945724E 1040112E 1040803E 1041624E 1042486E 1042718E 1042721E 1043803E 1046698E 1142619E 1143800E 1145679E 1146359E 114635E 1146663E 1242122E 1242714E 1242919E 1243536E 1243952E 1244528E 1244582E 1244993E 1245516E 1246169E 1246647E 1340192 1340541 1343411 1343710 1344029 1344298 1345423 1345645 1346655 1440435 1440679 1441180 1441180E 1441815 1442256 1442808 1443226 1443650
Other Recalls from Stryker Endoscopy
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2016 | Class II | PneumoSure XL High Flow Insufflator and Pneumo... | Mar 28, 2016 |
| Z-1831-2015 | Class II | SERFAS 90 degree Energy Probe, Part Number 279-... | Jun 3, 2015 |
| Z-0631-2015 | Class II | PKG, PENNINGTON FORCEPS, P/N 0250080242. Lapa... | Nov 17, 2014 |
| Z-0646-2015 | Class II | PKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250... | Nov 17, 2014 |
| Z-0679-2015 | Class II | PKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparo... | Nov 17, 2014 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.