Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product ...
FDA Device Recall #Z-1326-2016 — Class II — February 23, 2016
Recall Summary
| Recall Number | Z-1326-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 23, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Synthes (USA) Products LLC |
| Location | West Chester, PA |
| Product Type | Devices |
| Quantity | 500 units distributed in USA, and 35 units distributed OUS |
Product Description
Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.
Reason for Recall
Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of MA, VA, TN, GA, WA, IA, NJ, TN, PA, NY, CO, CT, VT, and the countries of Canada and Switzerland.
Lot / Code Information
Titanium Low Profile Neuro Screw, Self-Drilling, 3mm, part number # 400.833, Lot numbers: 9960359, 9962300, 9967875
Other Recalls from Synthes (USA) Products LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0228-2026 | Class II | LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM... | Sep 26, 2025 |
| Z-0229-2026 | Class II | LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM... | Sep 26, 2025 |
| Z-1916-2024 | Class II | STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/... | Apr 25, 2024 |
| Z-1239-2024 | Class II | TFNA Femoral Nail 9mm, right, 130¿, L 235mm-In... | Jan 24, 2024 |
| Z-1238-2024 | Class II | RAPIDSORB IPS Battery Pack, Sterile-Intended fo... | Jan 24, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.