Baxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a pharmaceutical comp...
FDA Device Recall #Z-3084-2024 — Class I — August 20, 2024
Recall Summary
| Recall Number | Z-3084-2024 |
| Classification | Class I — Serious risk |
| Date Initiated | August 20, 2024 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Baxter Healthcare Corporation |
| Location | Deerfield, IL |
| Product Type | Devices |
| Quantity | 5,828,540 units |
Product Description
Baxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a pharmaceutical compounding device
Reason for Recall
Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
Distribution Pattern
Worldwide distribution.
Lot / Code Information
UDI/DI 00085412475806, Lot Numbers: 803808 and lower
Other Recalls from Baxter Healthcare Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1573-2026 | Class II | Brand Name: Welch Allyn, Inc. Product Name: We... | Feb 18, 2026 |
| Z-1368-2026 | Class II | Brand Name: Spectrum IQ Product Name: Infusion... | Jan 13, 2026 |
| Z-1121-2026 | Class II | SIGMA Spectrum Infusion Pump, Model/Catalog Num... | Nov 28, 2025 |
| Z-0847-2026 | Class II | Welch Allyn INF BAG, Thigh 1-Tube, Model/Catalo... | Oct 28, 2025 |
| Z-0233-2026 | Class II | Oral Probe (Product code 02893-000), accessory ... | Sep 17, 2025 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.