Alere i flu, Alere i, Alere Influenza A & B; Part Number (PN) 425-024 The Alere" i Influenza A &...
FDA Device Recall #Z-2095-2015 — Class II — June 26, 2015
Recall Summary
| Recall Number | Z-2095-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 26, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Alere Scarborough, Inc. dba Binax, Inc. |
| Location | Scarborough, ME |
| Product Type | Devices |
| Quantity | 257 Kits |
Product Description
Alere i flu, Alere i, Alere Influenza A & B; Part Number (PN) 425-024 The Alere" i Influenza A & B assay, performed on the Alere" i Instrument, is a rapid molecular in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection and discrimination of influenza A and B viral RNA in nasal swabs from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the differential diagnosis of influenza A and B viral infections in humans, in conjunction with clinical and epidemiological risk factors. The assay is not intended to detect the presence of influenza C virus.
Reason for Recall
Risk of false negative results due to microbial growth in the Sample Receiver of one lot of the Alere i Influenza A & B kit
Distribution Pattern
Nationwide Distribution including AR CA CO CT FL GA IA IL IN KS KY LA MA MD ME MO MS NC NE NJ NY OH PA SC TX VA WA WY.
Lot / Code Information
Lot Number 0073853 (US distribution only)
Other Recalls from Alere Scarborough, Inc. dba Binax, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1677-2015 | Class II | Alere i Influenza A & B Product Usage: The ... | Apr 22, 2015 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.