The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement pro...
FDA Device Recall #Z-0953-2015 — Class II — November 10, 2014
Recall Summary
| Recall Number | Z-0953-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 10, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | DePuy Orthopaedics, Inc. |
| Location | Warsaw, IN |
| Product Type | Devices |
| Quantity | 7,944 |
Product Description
The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The instrument interfaces with several related ATTUNE instruments: keel punch, tibial tower, evaluation bullet, fixed bearing tibial impactor, rotating platform tibial impactor, and femoral impactor, for the purpose of impaction, extraction, and insertion applications at steps within the surgical procedure. Due to the frequency of its use, two impaction handles are supplied in each kit.
Reason for Recall
Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.
Distribution Pattern
Worldwide Distribution-US (nationwide) except AK and RI, including PR and the countries of Chile, Canada, Australia, Austria, Belgium, France, Hong Kong, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK
Lot / Code Information
CATALOG NO. : 2544-01-017 LOT NUMBER: All Lots Barcode GTIN: 10603295423324
Other Recalls from DePuy Orthopaedics, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1456-2024 | Class II | ATTUNE AFFIXIUM Cementless Fixed Bearing Knee w... | Mar 12, 2024 |
| Z-1264-2023 | Class II | Attune Revision Limb Preservation System (LPS) ... | Feb 15, 2023 |
| Z-1267-2023 | Class II | Attune Posterior (PS) Fixed Bearing (FB) Tibial... | Feb 15, 2023 |
| Z-1263-2023 | Class II | Attune Revision Limb Preservation System (LPS) ... | Feb 15, 2023 |
| Z-1266-2023 | Class II | Attune Revision Limb Preservation System (LPS) ... | Feb 15, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.