SmartPath to dStream for 1.5T, Model No. 781260, 782112

FDA Recall #Z-0812-2024 — Class III — December 29, 2023

Recall #Z-0812-2024 Date: December 29, 2023 Classification: Class III Status: Ongoing

Product Description

SmartPath to dStream for 1.5T, Model No. 781260, 782112

Reason for Recall

An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.

Recalling Firm

Philips North America — Cambridge, MA

Classification

Class III — Not likely to cause adverse health consequences.

Product Type

Devices

Product Quantity

7 US; 16 ROW total

Distribution

Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.

Code Information

Model No. 781260 UDI-DI 00884838095076 Serial Numbers 21020 9024 8657 21200 21223 18948 32131; Model No. 782112 UDI-DI 00884838098886 Serial Numbers 30089 20204

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated