cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clin...
FDA Device Recall #Z-0752-2021 — Class II — November 23, 2020
Recall Summary
| Recall Number | Z-0752-2021 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 23, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Roche Diagnostics Operations, Inc. |
| Location | Indianapolis, IN |
| Product Type | Devices |
| Quantity | 152 kits and 7 units |
Product Description
cobas pro integrated solutions -cobas c 503, Analyzer, chemistry (photometric discrete), for clinical use
Reason for Recall
During the subsequent pipetting (1st or 2nd measurement), the residual reagent can sporadically be transferred from the conus into the reaction cuvette and lead to inaccurately low results for the affected TDMs (Therapeutic Drug Monitoring).
Distribution Pattern
Domestic: IL, FL, GA, WV
Lot / Code Information
Catalog Number 08445605190 08445567190 08253153190 08058580190
Other Recalls from Roche Diagnostics Operations, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1476-2026 | Class II | cobas pro integrated solutions with cobas c 503... | Jan 20, 2026 |
| Z-1009-2026 | Class II | The Elecsys Anti-TSHR immunoassay is a three-st... | Nov 21, 2025 |
| Z-0489-2025 | Class II | cobas HCYS, Homocysteine Enzymatic Assay, Mater... | Oct 17, 2024 |
| Z-3187-2024 | Class II | Creatine Kinase (CK) used on cobas c 311/501/50... | Aug 15, 2024 |
| Z-0352-2024 | Class II | Tina-quant D-Dimer Test System, Fibrinogen/Fibr... | Oct 3, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.