GEM Premier 5000; Part No. 00055430004.

FDA Recall #Z-1537-2026 — Class II — January 20, 2026

Recall #Z-1537-2026 Date: January 20, 2026 Classification: Class II Status: Ongoing

Product Description

GEM Premier 5000; Part No. 00055430004.

Reason for Recall

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Recalling Firm

Instrumentation Laboratory — Bedford, MA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

253 units

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam.

Code Information

Part number: 00055430004; UDI: 08426950807711; Lots No.: 251111O, 251113U, 251117AD, 251118D, 251119D, 251120X, 251124Z, 251125S, 251201AI, 251201AJ, 251202P, 251203Q, 251208AA, 251212O, 251215J, 251216K, 251217I, 251217J, 251218V, 251219R, 251222T, 251222U, 251223E, 251223P, 251229AG, 251229X, 251230P.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated