Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory ...

FDA Recall #Z-0004-2025 — Class I — August 30, 2024

Recall #Z-0004-2025 Date: August 30, 2024 Classification: Class I Status: Ongoing

Product Description

Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.

Reason for Recall

Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step.

Recalling Firm

Mercury Enterprises, Inc. dba Mercury Medical — Clearwater, FL

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Product Quantity

1,300 units

Distribution

US distribution to MI, FL, SC, PA, IN. International distribution to Albania.

Code Information

Part Number: 1050805; UDI/DI: 10641043508053, 30641043508057; Lot Numbers: 2417650805, 2417750805. Part Number: 1050832; UDI/DI: 10641043508329, 30641043508323; Lot Numbers: 2423550832, 2417850832.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated