NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

FDA Recall #Z-0658-2022 — Class II — December 17, 2021

Recall #Z-0658-2022 Date: December 17, 2021 Classification: Class II Status: Terminated

Product Description

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Reason for Recall

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Recalling Firm

Nobel Biocare Usa Llc — Yorba Linda, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

702 pieces

Distribution

Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.

Code Information

Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated