VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 i...

FDA Device Recall #Z-2671-2025 - Class II, June 26, 2025

Recall Summary

Recall Number Z-2671-2025
Classification Class II - Moderate risk
Date Initiated June 26, 2025
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm bioMerieux, Inc.
Location Durham, NC
Product Type Devices
Quantity 235269 cards

Product Description

VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only

Reason for Recall

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

Distribution Pattern

International distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Macedonia, Malaysia/Brunei, Moldova, Netherlands, Pakistan, Poland, Portugal, Romania, Russia, Serbia, South Korea, Spain, Switzerland, Taiwan, Thailand, and Turkey.

Lot / Code Information

Material Number/Material Description/Batch Number: 423936 AST-N412 TEST KIT 20 CARDS 0143139204 423936 AST-N412 TEST KIT 20 CARDS 0143254504 423936 AST-N412 TEST KIT 20 CARDS 0143303504 424196 AST-N427 TEST KIT 20 CARDS 0542849204 424196 AST-N427 TEST KIT 20 CARDS 0543190104 424196 AST-N427 TEST KIT 20 CARDS 0543266504 424196 AST-N427 TEST KIT 20 CARDS 0542951404 424196 AST-N427 TEST KIT 20 CARDS 0543190104 424196 AST-N427 TEST KIT 20 CARDS 0543289404 424499 AST-N438 TEST KIT 20 CARDS 0703181504 424499 AST-N438 TEST KIT 20 CARDS 0703218104 424499 AST-N438 TEST KIT 20 CARDS 0703273104 424499 AST-N438 TEST KIT 20 CARDS 0703125104 424501 AST-N439 TEST KIT 20 CARDS 0723192104 424501 AST-N439 TEST KIT 20 CARDS 0723048404 424501 AST-N439 TEST KIT 20 CARDS 0722971504 424501 AST-N439 TEST KIT 20 CARDS 0723237504 424501 AST-N439 TEST KIT 20 CARDS 0722860204 424502 AST-N440 TEST KIT 20 CARDS 0733170404 424502 AST-N440 TEST KIT 20 CARDS 0733317504 424502 AST-N440 TEST KIT 20 CARDS 0733170404 424541 AST-N443 TEST KIT 20 CARDS 0773230404 424541 AST-N443 TEST KIT 20 CARDS 0773247404 424541 AST-N443 TEST KIT 20 CARDS 0772911104 424587 AST-N444 TEST KIT 20 CARDS 0802553214 424587 AST-N444 TEST KIT 20 CARDS 0802553204 424620 AST-N447 TEST KIT 20 CARDS 0853198104 424620 AST-N447 TEST KIT 20 CARDS 0853247204 424633 AST-N448 TEST KIT 20 CARDS 0862971204 424633 AST-N448 TEST KIT 20 CARDS 0863202504 424633 AST-N448 TEST KIT 20 CARDS 0863296504 424839 AST-N462 TEST KIT 20 CARDS No Lots 424840 AST-N463 TEST KIT 20 CARDS No Lots 424842 AST-N465 TEST KIT 20 CARDS No Lots 424891 AST-N475 TEST KIT 20 CARDS 1703184204 424891 AST-N475 TEST KIT 20 CARDS 1703225504 424934 AST-N476 TEST KIT 20 CARDS 1713192504 425019 AST-N477 TEST KIT 20 CARDS 1733184104 425019 AST-N477 TEST KIT 20 CARDS 1733205204 425019 AST-N477 TEST KIT 20 CARDS 1733205214 424197 AST-XN21 TEST KIT 20 CARDS 0493174204 424197 AST-XN21 TEST KIT 20 CARDS 0493310504 424199 AST-XN22 TEST KIT 20 CARDS 0552981204 424199 AST-XN22 TEST KIT 20 CARDS 0553156204 424351 AST-XN24 TEST KIT 20 CARDS 0653042204 424351 AST-XN24 TEST KIT 20 CARDS 0653217504 424394 AST-XN25 TEST KIT 20 CARDS 0662155104 424394 AST-XN25 TEST KIT 20 CARDS 0662553204 424586 AST-XN28 TEST KIT 20 CARDS 0793210504 424586 AST-XN28 TEST KIT 20 CARDS 0793265504 424586 AST-XN28 TEST KIT 20 CARDS 0793295504 424586 AST-XN28 TEST KIT 20 CARDS 0793210504 424604 AST-XN29 TEST KIT 20 CARDS 0822932204 424604 AST-XN29 TEST KIT 20 CARDS 0823262504 424810 AST-XN35 TEST KIT 20 CARDS 1323087204 424888 AST-XN37 TEST KIT 20 CARDS 1693079204 AST-XN25 424394 Australia AST-N412 423936 Thailand AST-XN21 424197 Turkey AST-XN22 424199 Iberia AST-N427 424196 Iberia AST-XN24 424351 Germany AST-N439 424501 Italy AST-N440 424502 Italy AST-XN28 424586 France AST-N443 424541 France, Spain & Portugal AST-N444 424587 Russia AST-N447 424620 Netherlands AST-N448 424633 Israel AST-XN35 424810 Poland AST-N462 424839 Taiwan AST-N463 424840 Taiwan AST-N465 424842 Taiwan AST-XN37 424888 Japan AST-N475 424891 Pakistan AST-N476 424934 Pakistan AST-N477 425019 India AST-N438 424499 Italy AST-XN29 424604 Japan

Other Recalls from bioMerieux, Inc.

Recall # Classification Product Date
Z-1564-2022 Class II NUCLISENS easyMAG Magnetic Silica, REF 280133, ... Jul 20, 2022
Z-1640-2022 Class II VITEK 2 / MYLA VITEK 2 Software versions 8.0... Jul 13, 2022
Z-1059-2022 Class II VITEK 2 Systems and VITEK 2 with MYLA. Mar 29, 2022
Z-0742-2022 Class II VIDAS CMV IgM, 30 tests, REF 30205-01, IVD, Rx. Feb 9, 2022
Z-0243-2022 Class II API 50 CH, Model 50300 Sep 30, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.