BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infu...
FDA Device Recall #Z-1231-2025 — Class II — January 23, 2025
Recall Summary
| Recall Number | Z-1231-2025 |
| Classification | Class II — Moderate risk |
| Date Initiated | January 23, 2025 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | CareFusion 303, Inc. |
| Location | San Diego, CA |
| Product Type | Devices |
| Quantity | 242,098 systems |
Product Description
BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration
Reason for Recall
Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.
Distribution Pattern
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Guantanamo Bay, Cuba O.U.S.: Canada
Lot / Code Information
Catalog Number: 8100 UDI-DI codes: 10885403810046 10885403517723 Software Version: 12.1.2 Serial/Lot Numbers: Pending
Other Recalls from CareFusion 303, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1645-2026 | Class II | BD Alaris" System with Guardrails" Suite MX wit... | Feb 11, 2026 |
| Z-1028-2026 | Class II | BD Pyxis ES Enterprise Server Catalog Numbers 1... | Dec 3, 2025 |
| Z-0962-2026 | Class II | BD Pyxis MedStation ES, REF: 323, BD Pyxis M... | Nov 20, 2025 |
| Z-1217-2026 | Class II | BD Alaris" Pump Module model 8100 UDI-DI code:... | Nov 6, 2025 |
| Z-0430-2026 | Class I | BD Alaris infusion Pump Module 8100, REF: 8100;... | Oct 17, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.