Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems ...

FDA Device Recall #Z-2656-2016 — Class II — July 20, 2016

Recall Summary

Recall Number Z-2656-2016
Classification Class II — Moderate risk
Date Initiated July 20, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Philips Medical Systems (Cleveland) Inc
Location Cleveland, OH
Product Type Devices
Quantity 50

Product Description

Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Reason for Recall

After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

Distribution Pattern

US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY

Lot / Code Information

728240 -- Brilliance CT 16 Power 3100 5043 5047 5280 5345 5390 5413 5430 5495 5499 5512 6007 6009 6013 6032 6041 6045 6049 6051 6058 6064 6066 6070 6071 6077 6080 6081 6084 6089 6095 6100 6106 6109 6113 6115 6116 6123 6124 6131 6132 6134 6141 6144 6155 6161 6162 5476 6098 6120 6067

Other Recalls from Philips Medical Systems (Cleveland) Inc

Recall # Classification Product Date
Z-2570-2025 Class II Pinnacle Radiation Therapy Planning System: ... Aug 5, 2025
Z-2239-2025 Class II Pinnacle 3 with TumorLOC, (870258) Radiation Th... Jul 17, 2025
Z-1744-2022 Class II Pinnacle3 Radiation Therapy Planning System, Mo... Aug 30, 2022
Z-0213-2020 Class I Forte Gamma Camera System is intended to produc... Sep 19, 2019
Z-0217-2020 Class I Forte Gamma Camera System is intended to produc... Sep 19, 2019

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.