AirFit F30 Full Face Mask and User Guide

FDA Device Recall #Z-0541-2024 — Class I — November 20, 2023

Recall Summary

Recall Number Z-0541-2024
Classification Class I — Serious risk
Date Initiated November 20, 2023
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm ResMed Ltd.
Location Bella Vista, N/A
Product Type Devices
Quantity 2,660,776

Product Description

AirFit F30 Full Face Mask and User Guide

Reason for Recall

Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact

Distribution Pattern

US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS, IT, JE, JO, JP, KE, KG, KR, KW, LB, LK, LT, LV, LY, MA, MC, MM, MQ, MT, MU, MV, MX, MY, NA, NC, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PM, PT, QA, RE, RO, RS, RU, SA, SD, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, WF, YT, ZA

Lot / Code Information

All mask lots used with User Guide 638234/version 2020-07, and prior. UDI-DI/GTIN: 619498641048, 619498641000, 619498641017, 619498641406, 619498641055, 619498641116, 619498641277, 619498641109, 619498641154, 619498641260, 619498641147

Other Recalls from ResMed Ltd.

Recall # Classification Product Date
Z-0538-2024 Class I AirTouch F20 Full Face Mask and User Guide Nov 20, 2023
Z-0540-2024 Class I AirTouch N20 Nasal Mask and User Guide Nov 20, 2023
Z-0536-2024 Class I AirFit N10 Nasal Masks and User Guide Nov 20, 2023
Z-0537-2024 Class I AirFit F20 Full Face Mask and User Guide Nov 20, 2023
Z-0542-2024 Class I AirFit F30i Full Face Mask and User Guide Nov 20, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.