SIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic...
FDA Device Recall #Z-0876-2020 - Class II, December 23, 2019
Recall Summary
| Recall Number | Z-0876-2020 |
| Classification | Class II - Moderate risk |
| Date Initiated | December 23, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Healthcare, LLC |
| Location | Waukesha, WI |
| Product Type | Devices |
| Quantity | 62 (30 US, 32 OUS) total devices |
Product Description
SIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Reason for Recall
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Distribution Pattern
Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Lot / Code Information
Mfg. Lot or Serial # System ID To be provided GON4833100 01264 860536MR3T KZTBL1700064TJ 860871MR3T KZTBL1800069TJ RADNETMR1700 KZTBL1800022TJ 508872PMR 01302 RADNETMR1708 KZTBL1900130TJ 301698MR1 KZTBL1900071TJ 704333PIO KZTBL1900110TJ 631751MR KZTBL1900107TJ RADNETMR1707 KZTBL1900121TJ 803936LMCMR2 033-4-1532 281HMI750W To be provided 83199160995 To be provided M24313812 To be provided M2850013 To be provided TP0013MR01 To be provided PNUE04MR02 To be provided TP0008MR07 To be provided 083027823408018 To be provided 083027823888617 To be provided GON4826677 To be provided GON4762453 To be provided B5142305 To be provided ZC0002 To be provided ZC0034 To be provided ZC0001 To be provided ZC0306 To be provided GON4795546 To be provided 222768MR01 To be provided 34340MRS02 To be provided 10627MRS02
Other Recalls from GE Healthcare, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2022 | Class II | (1) GE Centricity Universal Viewer 6.0, Model 2... | Jul 15, 2022 |
| Z-1430-2022 | Class II | SIGNA Premier magnetic resonance scanner, model... | Jun 24, 2022 |
| Z-1466-2022 | Class II | Centricity PACS Software Version 7.0 SP0.0.4.7 | Jun 14, 2022 |
| Z-1333-2022 | Class II | GE Centricity Universal Viewer Zero Footprint. ... | Jun 8, 2022 |
| Z-1511-2022 | Class II | Service manuals for the CT, PET, or NM table as... | Jun 2, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.