"***PressFT***2.6 w/ Two #1 (4 metric) Hi-Fi***Sutures***NP262***2009-02***CONMED LINVATEC 11311 ...

FDA Device Recall #Z-0063-2013 — Class II — June 26, 2012

Recall Summary

Recall Number Z-0063-2013
Classification Class II — Moderate risk
Date Initiated June 26, 2012
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Linvatec Corp. dba ConMed Linvatec
Location Largo, FL
Product Type Devices
Quantity 879 units for all products in recall event

Product Description

"***PressFT***2.6 w/ Two #1 (4 metric) Hi-Fi***Sutures***NP262***2009-02***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FL 33773-4908 USA*** To reattach soft tissue to bone in orthopedic surgical procedures

Reason for Recall

Linvatec Corporation dba ConMed Linvatec in Largo, FL is recalling PressFT HI-FI Sutures (items NP211, NP212, NP261, NP262, NP211H, NP261H-all lot numbers) due to the potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.

Distribution Pattern

Worldwide Distribution - USA including AZ, CA, CO, FL, GA, LA, MA, ME, MI, MN, NJ, NM, NY, PA, TX, UT, and WA. Internationally to BE, CA, ES, MY, NZ and PL

Lot / Code Information

Item number NP262

Other Recalls from Linvatec Corp. dba ConMed Linvatec

Recall # Classification Product Date
Z-1680-2013 Class II ***REF C7120***APEX Arthroscopy Tubing Set. ... May 30, 2013
Z-1681-2013 Class II ***REF C7122***QTY 1***APEX Arthroscopy Tubing ... May 30, 2013
Z-1340-2013 Class II Ref: VP1600 HD Still Capture System ConMed Linv... Mar 18, 2013
Z-1619-2013 Class II GENESYS Cross FT Suture Anchor with Two #2 (5 m... Mar 6, 2013
Z-0916-2013 Class III D3000 Advantage Drive System, 115v Product ... Jan 24, 2013

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.