AXIOM Artis angiographic x-ray system, Model Numbers: 7412807, 7413078, 7728392 AXIOM Artis i...

FDA Device Recall #Z-1150-2018 — Class II — January 5, 2018

Recall Summary

Recall Number Z-1150-2018
Classification Class II — Moderate risk
Date Initiated January 5, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Medical Solutions USA, Inc
Location Malvern, PA
Product Type Devices
Quantity 22

Product Description

AXIOM Artis angiographic x-ray system, Model Numbers: 7412807, 7413078, 7728392 AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason for Recall

Certain systems may have leaking of the plastic coolant couplings, the liquid can enter into the flat panel detector leading to a failure of the detector. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.

Distribution Pattern

US Distribution in the following U.S. states: AZ, CA, FL, IA, IL, LA, MI, NC, NJ, OH, OK, and TX.

Lot / Code Information

Serial Numbers: 35061 35054 34999 35098 35059 35064 35073 35071 35057 35075 35083 35018 35097 35108 35046 35048 35050 46004 35045 35004 44011 35095

Other Recalls from Siemens Medical Solutions USA, Inc

Recall # Classification Product Date
Z-1175-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1174-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1241-2026 Class II NAEOTOM Alpha.Prime Software applications: s... Dec 19, 2025
Z-1240-2026 Class II NAEOTOM Alpha Software applications: syngo.C... Dec 19, 2025
Z-1239-2026 Class II SOMATOM X.ceed Software applications: syngo.... Dec 19, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.