Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Co...

FDA Device Recall #Z-1793-2012 — Class II — February 21, 2012

Recall Summary

Recall Number Z-1793-2012
Classification Class II — Moderate risk
Date Initiated February 21, 2012
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Aesculap Implant Systems
Location Center Valley, PA
Product Type Devices
Quantity 8

Product Description

Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplastics or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture

Reason for Recall

During an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of the extension stem.

Distribution Pattern

Nationwide Distribution including PA and TX.

Lot / Code Information

NR068K, TIBIA PLATEAU, FIXED, T0; NR070K, TIBIA PLATEAU, FIXED, T0+; NR072K, TIBIA PLATEAU, FIXED, T1; NR072K, TIBIA PLATEAU, FIXED, T1+; NR073K, TIBIA PLATEAU, FIXED, T2; NR074K, TIBIA PLATEAU, FIXED, T2+; NR075K, TIBIA PLATEAU, FIXED, T3; NR076K, TIBIA PLATEAU, FIXED, T3+; NR077K, TIBIA PLATEAU, FIXED, T4; NR078K, TIBIA PLATEAU, FIXED, T4+; NR079K, TIBIA PLATEAU, FIXED, T5

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.