Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (E...

FDA Device Recall #Z-1493-2014 — Class II — March 26, 2014

Recall Summary

Recall Number Z-1493-2014
Classification Class II — Moderate risk
Date Initiated March 26, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Stryker Instruments Div. of Stryker Corporation
Location Portage, MI
Product Type Devices
Quantity 35,310 unexpired units

Product Description

Stryker PenAdapt"; Rx Only, Sterile. This device is an accessory to an electrosurgical unit (ESU). The PenAdapt is used to remove smoke, particles, and body and casual fluids from the point of surgical activity during surgical procedures that use ESU for cutting and cauterizing.

Reason for Recall

During packaging verification testing, a failure occurred due to cuts in the packaging potentially causing a breach in the sterile barrier. Additional testing identified microscopic aluminum metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches that were introduced during the manufacturing process and were sterilized within the

Distribution Pattern

Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV,and WY, and the countries of Austria, Canada, Sweden, England, UAE, Netherlands, Spain, and Hong Kong.

Lot / Code Information

Part Number 0702-045-027

Other Recalls from Stryker Instruments Div. of Stryker C...

Recall # Classification Product Date
Z-1452-2022 Class II SurgiCount+ System Application, Software Versio... Jun 15, 2022
Z-0431-2021 Class II Stryker Zyphr Disposable Cranial Perforator, La... Oct 16, 2020
Z-1807-2020 Class II Neptune E-SEP 165mm Blade Electrode, Catalog Nu... Mar 26, 2020
Z-1402-2022 Class II Triton Canister System (finished part numbers F... Jan 22, 2020
Z-1097-2020 Class II SafeAir Telescopic Smoke Evacuation Pencil, Pr... Jan 13, 2020

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.