BD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for collection and transpor...
FDA Device Recall #Z-0392-2018 — Class II — July 18, 2017
Recall Summary
| Recall Number | Z-0392-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 18, 2017 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Becton Dickinson & Co. |
| Location | Sparks, MD |
| Product Type | Devices |
| Quantity | N/A |
Product Description
BD MAXTM Vaginal Panel. Catalog Number(s): 443712 443710 443711 (RUO)for collection and transport of vaginal and endocervical swab specimens, in addition to the preservation of urine from males and females.
Reason for Recall
BD recently notified customers in July 2017 regarding an increased rate of unresolved (UNR) and indeterminate (IND) results for the BD MAXTM Vaginal Panel, experienced by some customers. A non-reportable result may lead to repeat of the test and/or re-collection of a specimen.
Distribution Pattern
US and OUS.
Lot / Code Information
UDI: 30382904433766. All Lot numbers: Not Distributed in US Not Distributed in US
Other Recalls from Becton Dickinson & Co.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0484-2026 | Class II | BD Kiestra" Isolate Suspension Cuvette Array; ... | Oct 23, 2025 |
| Z-0235-2026 | Class II | BD EpiCenter Microbiology Data Management Syste... | Sep 23, 2025 |
| Z-0236-2026 | Class II | BD BACTEC Blood Culture System, labeled as the ... | Sep 23, 2025 |
| Z-0240-2026 | Class II | BD Veritor Connect Software, labeled as the fol... | Sep 23, 2025 |
| Z-0237-2026 | Class II | BD BACTEC Blood Culture System, labeled as the ... | Sep 23, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.