WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2...
FDA Device Recall #Z-2878-2018 — Class I — May 7, 2018
Recall Summary
| Recall Number | Z-2878-2018 |
| Classification | Class I — Serious risk |
| Date Initiated | May 7, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Biosense Webster, Inc. |
| Location | Irwindale, CA |
| Product Type | Devices |
| Quantity | 25 units |
Product Description
WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.
Reason for Recall
When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.
Distribution Pattern
No U.S. Distribution. International Distribution only.
Lot / Code Information
All lots.
Other Recalls from Biosense Webster, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1205-2025 | Class I | VARIPULSE Bi-Directional Ablation Catheter REF... | Jan 5, 2025 |
| Z-1290-2024 | Class II | CARTO VIZIGO Bi-Directional Guiding Sheath, RE... | Feb 16, 2024 |
| Z-1319-2023 | Class II | Biosense Webster Carto Vizigo Bi-Directional Gu... | Feb 21, 2023 |
| Z-1320-2023 | Class II | Biosense Webster Carto Vizigo Bi-Directional Gu... | Feb 21, 2023 |
| Z-1677-2022 | Class II | CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath... | Jun 29, 2022 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.