DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagno...

FDA Device Recall #Z-0107-2024 — Class II — September 12, 2023

Recall Summary

Recall Number Z-0107-2024
Classification Class II — Moderate risk
Date Initiated September 12, 2023
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Beckman Coulter, Inc.
Location Chaska, MN
Product Type Devices
Quantity 36 units

Product Description

DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic

Reason for Recall

The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.

Distribution Pattern

Worldwide - US Nationwide distribution in the state of Florida, and the countries of Croatia, Czechia, France, Germany, Ireland, Italy, Netherlands, New Zealand, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.

Lot / Code Information

UDI/DI 15099590732103, All DxI 9000 Access Immunoassay Analyzer running software version 1.15 or prior

Other Recalls from Beckman Coulter, Inc.

Recall # Classification Product Date
Z-1235-2026 Class II CellMek SPS Sample Preparation System, REF: C44... Dec 23, 2025
Z-1025-2026 Class II Beckman Coulter UniCel Dxl 600, Part numbers: ... Nov 26, 2025
Z-1026-2026 Class II Beckman Coulter UniCel Dxl 800, Part numbers: ... Nov 26, 2025
Z-1221-2026 Class II COULTER DxH Diluent, REF 628017 COULTER DxH EC... Nov 10, 2025
Z-0889-2026 Class II Access 2 Reaction Vessels, individual, disposab... Nov 7, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.