Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to meas...

FDA Device Recall #Z-0123-2017 — Class II — August 2, 2016

Recall Summary

Recall Number Z-0123-2017
Classification Class II — Moderate risk
Date Initiated August 2, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Randox Laboratories Ltd.
Location Crumlin (North), N/A
Product Type Devices
Quantity 18

Product Description

Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.

Reason for Recall

According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.

Distribution Pattern

US Nationwide Distribution

Lot / Code Information

Catalogue number: LI3837, LI7979, LI8050, LI8361

Other Recalls from Randox Laboratories Ltd.

Recall # Classification Product Date
Z-2425-2024 Class II RX Series CYSC- Cystatin C Reagent -Intended fo... Jun 7, 2024
Z-1786-2024 Class III Evidence MultiSTAT with software v 3.7-Analyzer... Apr 5, 2024
Z-1759-2024 Class II Clinical Chemistry Calibration Serum (Catalog n... Mar 5, 2024
Z-1328-2024 Class II Liquid Protein Calibrators. C3 and Haptoglobin ... Mar 1, 2024
Z-1369-2024 Class II Cystatin C Calibrator - In vitro diagnostic (IV... Feb 14, 2024

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.