MRI system: Vantage Elan, Model: MRT-2020

FDA Device Recall #Z-1249-2024 — Class II — January 29, 2024

Recall Summary

Recall Number Z-1249-2024
Classification Class II — Moderate risk
Date Initiated January 29, 2024
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Canon Medical System, USA, INC.
Location Tustin, CA
Product Type Devices
Quantity 1

Product Description

MRI system: Vantage Elan, Model: MRT-2020

Reason for Recall

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.

Lot / Code Information

UDI-DI: 04987670101632. Serial Number: S1F2052753

Other Recalls from Canon Medical System, USA, INC.

Recall # Classification Product Date
Z-1018-2026 Class II VANTAGE TITAN 3T Model MRT-3010/MEXL-3010 Dec 19, 2025
Z-1177-2026 Class II Alphenix INFX-8000F, interventional fluoroscopi... Dec 19, 2025
Z-1017-2026 Class II VANTAGE GALAN 3T Model MRT-3020/MEXL-3020 Dec 19, 2025
Z-1178-2026 Class II Alphenix INFX-8000H, interventional fluoroscopi... Dec 19, 2025
Z-1179-2026 Class II Alphenix INFX-8000V, interventional fluoroscopi... Dec 19, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.