ACCU-CHECK FastClix Lancing Device. It is intended for use in conjunction with a blood glucose mo...

FDA Device Recall #Z-2109-2013 — Class II — August 19, 2013

Recall Summary

Recall Number Z-2109-2013
Classification Class II — Moderate risk
Date Initiated August 19, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Roche Diagnostics Operations, Inc.
Location Indianapolis, IN
Product Type Devices
Quantity 36,332

Product Description

ACCU-CHECK FastClix Lancing Device. It is intended for use in conjunction with a blood glucose monitoring system.

Reason for Recall

Internal Roche inspections have revealed that after the lancet is used for a finger stick, the needle may not fully retract back within the lancet drum or back into the cap. This allows the needle to protrude inside the cap or outside of the cap of the ACCU-CHEK FastClix lancing device.

Distribution Pattern

Nationwide Distribution

Lot / Code Information

Lot Number(s): GDA 048, GDA050, GDA 051, GDA 063

Other Recalls from Roche Diagnostics Operations, Inc.

Recall # Classification Product Date
Z-1476-2026 Class II cobas pro integrated solutions with cobas c 503... Jan 20, 2026
Z-1009-2026 Class II The Elecsys Anti-TSHR immunoassay is a three-st... Nov 21, 2025
Z-0489-2025 Class II cobas HCYS, Homocysteine Enzymatic Assay, Mater... Oct 17, 2024
Z-3187-2024 Class II Creatine Kinase (CK) used on cobas c 311/501/50... Aug 15, 2024
Z-0352-2024 Class II Tina-quant D-Dimer Test System, Fibrinogen/Fibr... Oct 3, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.