Giraffe Spot PT Ohmeda Medical Spot PT Lite Phototherapy System- The Spot PT Lite Phototherapy S...
FDA Device Recall #Z-1503-2017 — Class II — November 18, 2016
Recall Summary
| Recall Number | Z-1503-2017 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 18, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Ohmeda Medical |
| Location | Laurel, MD |
| Product Type | Devices |
| Quantity | 5,993 (5,080 US; 913 OUS) |
Product Description
Giraffe Spot PT Ohmeda Medical Spot PT Lite Phototherapy System- The Spot PT Lite Phototherapy System provides light therapy for the treatment of hyperbilirubinemia, commonly known as neonatal jaundice, during the newborn period in the hospital
Reason for Recall
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Distribution Pattern
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
Lot / Code Information
See Attached Spreadsheat
Other Recalls from Ohmeda Medical
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0041-2020 | Class II | Giraffe Omnibed - Product Usage: The OmniBed is... | Oct 5, 2018 |
| Z-1181-2018 | Class II | Rotating IV Pole Rotating IV Pole falls un... | Dec 1, 2017 |
| Z-1502-2017 | Class II | Giraffe Incubator The Giraffe Incubator is an ... | Nov 18, 2016 |
| Z-1504-2017 | Class II | Giraffe OmniBed Ohmeda Medical Giraffe OmniBed... | Nov 18, 2016 |
| Z-1511-2017 | Class II | IWS Infant Warmer System-Infant radiant warmer... | Nov 18, 2016 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.