The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction wit...

FDA Recall #Z-1622-2025 — Class II — March 20, 2025

Recall #Z-1622-2025 Date: March 20, 2025 Classification: Class II Status: Ongoing

Product Description

The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting Cannula has four lumens which house the endoscope, C-Ring, distal lens washer tube, and Harvesting Tool for cutting and cauterizing vessel branches. The Harvesting Tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning the device, cutting, and cauterization are performed under endoscopic visualization. The device is intended for specific use with the VASOVIEW HEMOPRO Power Supply, VASOVIEW HEMOPRO 2 extension cable, and an adapter cable.

Reason for Recall

There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.

Recalling Firm

Maquet Cardiovascular, LLC — Wayne, NJ

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

22,244 units

Distribution

Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.

Code Information

Model No. VH-4000 (Hemopro 2), VH-4001 (Hemopro 2 w/ Vasoshield); UDI-DI: 00607567700406 (Hemopro 2), 00607567700451 (Hemopro 2 w/ Vasoshield); Lot/Serial No. 3000436447, 3000437191, 3000439557, 3000439600, 3000439623, 3000439922, 3000440187, 3000440220, 3000440283, 3000440754, 3000440957, 3000440981, 3000440982, 3000441144, 3000441319, 3000441541, 3000441774, 3000442133, 3000442263, 3000442323, 3000442627, 3000442871, 3000442953, 3000442955, 3000443198, 3000443216, 3000443545, 3000443632, 3000443851, 3000444131, 3000444151, 3000444484, 3000445702, 3000445704, 3000445941, 3000446003, 3000446365, 3000446489, 3000446709, 3000446774, 3000447195, 3000447535, 3000447597, 3000447845, 3000448208, 3000448471, 3000448761, 3000449047, 3000449170, 3000449267, 3000450191, 3000450967.

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated