MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Ba...

FDA Recall #Z-1013-2026 — Class I — November 26, 2025

Recall #Z-1013-2026 Date: November 26, 2025 Classification: Class I Status: Ongoing

Product Description

MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.

Reason for Recall

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Recalling Firm

Medline Industries, LP — Northfield, IL

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Product Quantity

138412 units

Distribution

US, CA, PA, GUAM, VI

Code Information

1) REF MDR107003E, UDI/DI 40080196321002, All lots; 2) REF MDR107003E-4, UDI/DI 40080196321002, All lots

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated