GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and mu...
FDA Device Recall #Z-0841-2015 — Class II — September 15, 2014
Recall Summary
| Recall Number | Z-0841-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | September 15, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE OEC Medical Systems, Inc |
| Location | Salt Lake City, UT |
| Product Type | Devices |
| Quantity | 61 units installed in US |
Product Description
GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Reason for Recall
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Distribution Pattern
Worldwide Distribution
Lot / Code Information
GE OEC Brivo 865 Plus Serial Numbers: B4S13011 B4S13012 B4S13016 B4S13018 B4S13020 B4S13029 B4S13031 B4S13035 B4S13038 B4S13043 B4S13054 B4S13056 B4S13059- B4S13061 B4S14002- B4S14004 B4S14007 B4S14015 B4S14016 B4S14022 B4S14022 B4S1403 3 B4S1403 4
Other Recalls from GE OEC Medical Systems, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0703-2025 | Class II | OEC 3D REF Image-Intensified Fluoroscopic X... | Nov 21, 2024 |
| Z-0705-2025 | Class II | OEC 9900 REF Image-Intensified Fluoroscopic... | Nov 21, 2024 |
| Z-0704-2025 | Class II | OEC Elite REF Image-Intensified Fluoroscopi... | Nov 21, 2024 |
| Z-0706-2025 | Class II | OEC 9800 REF Image-Intensified Fluoroscopic... | Nov 21, 2024 |
| Z-2279-2023 | Class II | Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TO... | May 15, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.