Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The ...

FDA Device Recall #Z-0574-2015 — Class II — August 26, 2013

Recall Summary

Recall Number Z-0574-2015
Classification Class II — Moderate risk
Date Initiated August 26, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Zimmer, Inc.
Location Warsaw, IN
Product Type Devices
Quantity 14 devices

Product Description

Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2 Polyaxial Spinal System, used for posterior lumbar fusion procedures.

Reason for Recall

Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.

Distribution Pattern

US nationwide distribution.

Lot / Code Information

Lot: L75066

Other Recalls from Zimmer, Inc.

Recall # Classification Product Date
Z-0981-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0979-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0973-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0977-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0967-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.