ATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE SCREW MODULE, REF 176-504, QTY: 1 EA, Rx ...
FDA Device Recall #Z-0037-2013 — Class II — August 30, 2012
Recall Summary
| Recall Number | Z-0037-2013 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 30, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Medtronic Sofamor Danek USA Inc |
| Location | Memphis, TN |
| Product Type | Devices |
| Quantity | 1014 units |
Product Description
ATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE SCREW MODULE, REF 176-504, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.
Reason for Recall
Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.
Distribution Pattern
Worldwide Distribution-USA (nationwide) and the countries of India, Netherlands, Hong Kong, S. L Korea, Singapore, Canada, Australia, Japan, and Mexico.
Lot / Code Information
All lots except those starting with SY
Other Recalls from Medtronic Sofamor Danek USA Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0580-2026 | Class II | Catalyft PL & PL40 Expandable Interbody System ... | Sep 16, 2025 |
| Z-0162-2024 | Class II | Grafton DBM (Flex), Model Numbers: T42100, Demi... | Sep 26, 2023 |
| Z-0165-2024 | Class II | Grafton DBM Orthoblend, Model Numbers: a) T4412... | Sep 26, 2023 |
| Z-0166-2024 | Class II | Grafton Plus DBM Paste, Model Numbers: a) T4500... | Sep 26, 2023 |
| Z-0167-2024 | Class II | Grafton DBM Putty, Model Numbers: a) T43102, b)... | Sep 26, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.