Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter...

FDA Recall #Z-0853-2015 — Class II — December 4, 2014

Recall #Z-0853-2015 Date: December 4, 2014 Classification: Class II Status: Terminated

Product Description

Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.

Reason for Recall

The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi

Recalling Firm

Zimmer, Inc. — Warsaw, IN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

83 units

Distribution

Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.

Code Information

Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated