AXIOM Artis dFA, Model Number 7555373

FDA Device Recall #Z-2024-2019 — Class II — June 12, 2019

Recall Summary

Recall Number Z-2024-2019
Classification Class II — Moderate risk
Date Initiated June 12, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Medical Solutions USA, Inc
Location Malvern, PA
Product Type Devices
Quantity 39

Product Description

AXIOM Artis dFA, Model Number 7555373

Reason for Recall

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Distribution Pattern

US Nationwide Distribution

Lot / Code Information

Serial Numbers: 50063 50059 50006 50064 50189 50044 50074 50101 51719 50062 50065 50087 50167 50039 50084 51706 51707 50048 50161 50093 50206 50057 50081 50133 50040 50016 50120 50191 51701 50027 51700 50139 50004 50010 50024 50142 50079 50132 50050

Other Recalls from Siemens Medical Solutions USA, Inc

Recall # Classification Product Date
Z-1175-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1174-2026 Class II Interventional Fluoroscopic X-Ray System: LUMIN... Dec 29, 2025
Z-1241-2026 Class II NAEOTOM Alpha.Prime Software applications: s... Dec 19, 2025
Z-1240-2026 Class II NAEOTOM Alpha Software applications: syngo.C... Dec 19, 2025
Z-1239-2026 Class II SOMATOM X.ceed Software applications: syngo.... Dec 19, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.