GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/...
FDA Device Recall #Z-0241-2023 — Class II — October 17, 2022
Recall Summary
| Recall Number | Z-0241-2023 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 17, 2022 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Steris Corporation |
| Location | Mentor, OH |
| Product Type | Devices |
| Quantity | 5 units |
Product Description
GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001
Reason for Recall
Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result
Distribution Pattern
US Nationwide distribution in the states of Arizona, Illinois, Tennessee, Virginia.
Lot / Code Information
UDI: 00724995180256 Serial Numbers: 0032-0822-003, 0032-0622-009, 0032-0922-001, 0032-0822-001, and 0032-0082-002
Other Recalls from Steris Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0910-2026 | Class II | AMSCO 7053HP Washer/Disinfector Model/Catalog... | Nov 19, 2025 |
| Z-0909-2026 | Class II | AMSCO 7052HP Washer/Disinfector Model/Catalog ... | Nov 19, 2025 |
| Z-2178-2025 | Class II | Dual Flat Panel Monitor (DFPM) accessories for ... | Jun 23, 2025 |
| Z-2177-2025 | Class II | Dual Flat Panel Monitor (DFPM) accessories for ... | Jun 23, 2025 |
| Z-1361-2026 | Class II | Barco MNA with the HexaVue IP Integration Syste... | May 23, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.