The Tina-Quant Myoglobin Gen. 2 Tests system is an immuno-turbidimetric assay for the quantitativ...
FDA Device Recall #Z-2445-2019 — Class II — July 2, 2019
Recall Summary
| Recall Number | Z-2445-2019 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 2, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Roche Diagnostics Operations, Inc. |
| Location | Indianapolis, IN |
| Product Type | Devices |
| Quantity | 4,309 (cobas c311-340 units;cobas c501-3,517 units;cobas c502 452 units |
Product Description
The Tina-Quant Myoglobin Gen. 2 Tests system is an immuno-turbidimetric assay for the quantitative in vitro determination of myoglobin in human serum and plasma on Roche automated clinical chemistry analyzers
Reason for Recall
The firm conducted internal investigations that confirmed customer complaints of quality control (QC) imprecision and calibration failures for the Tina-quant Myoglobin Gen.2 assay (catalog number 04580010190), lot number 349860 on the cobas c311 analyzer and cobas c501 and c502 modules due to a drop in signal in the reaction kinetics. This signal drop was traced to an adjacent cell mixing effect. Sample recovery is decreased as a consequence of the disturbance in reaction kinetics. The investigation found discrepancies of up to -30%. Calibration, QC, and patient samples can be affected. The following issues were reported: - Calibration failures; Dup.E error due to signal drop in reaction kinetics - Sporadic quality control (QC) imprecision. As a long-term solution, the firm will implement required extra wash cycles (EWCs) into the respective Special Wash Requirements method sheets for the cobas c 311 analyzer and cobas c 501 and 502 modules, and into the cobas e-file for the cobas c 502 module. For the use of the reagent with the cobas c501 and c502, Roche will also direct users to run the Tina-Quant Myoglobin gen. 2 assay independently from other tests, in a batch mode status, with a maximum of 39 samples per run. Additionally, a change in the application settings of the Tina-quant Myoglobin Gen.2 assay on the cobas c 501 and 502 modules will be implemented so that adjacent cuvettes are no longer used in order to prevent adjacent cell mixing effects. The application settings change will be communicated at a later date.
Distribution Pattern
Nationwide
Lot / Code Information
Lot: 349860 and onward
Other Recalls from Roche Diagnostics Operations, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1476-2026 | Class II | cobas pro integrated solutions with cobas c 503... | Jan 20, 2026 |
| Z-1009-2026 | Class II | The Elecsys Anti-TSHR immunoassay is a three-st... | Nov 21, 2025 |
| Z-0489-2025 | Class II | cobas HCYS, Homocysteine Enzymatic Assay, Mater... | Oct 17, 2024 |
| Z-3187-2024 | Class II | Creatine Kinase (CK) used on cobas c 311/501/50... | Aug 15, 2024 |
| Z-0352-2024 | Class II | Tina-quant D-Dimer Test System, Fibrinogen/Fibr... | Oct 3, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.