Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J
FDA Device Recall #Z-2413-2020 — Class I — May 20, 2020
Recall Summary
| Recall Number | Z-2413-2020 |
| Classification | Class I — Serious risk |
| Date Initiated | May 20, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Arrow International Inc |
| Location | Reading, PA |
| Product Type | Devices |
| Quantity | 2123 total |
Product Description
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J
Reason for Recall
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Distribution Pattern
Worldwide distribution.
Lot / Code Information
Serial Number Ranges: AUTOCAT2 WAVE JAPANESE IAP-0500J 140866W through 140867W 150117W through 150118W 150120W 150201W 150305W 150307W through 150309W 150623W through 150624W 150713W through 150716W 151114W 151209W 151213W 151219W 160503W 160511W through 160515W 160517W 160635W through 160636W 160640W through 160641W 160731W through 160733W 171105W 171117W 171119W through 171120W 171201W 180120W 180301W
Other Recalls from Arrow International Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0231-2022 | Class I | Arrow-Trerotola Over-The-Wire PTD Kit Percutane... | Sep 20, 2021 |
| Z-0232-2022 | Class I | Arrow-Trerotola Over-The-Wire PTD Kit Percutane... | Sep 20, 2021 |
| Z-0234-2022 | Class I | Arrow-Trerotola Over-The-Wire PTD Kit Percutane... | Sep 20, 2021 |
| Z-0233-2022 | Class I | Arrow-Trerotola Over-The-Wire PTD Kit Percutane... | Sep 20, 2021 |
| Z-2071-2021 | Class II | Arrow Temporary Pacing Catheter/Introducer Kit ... | May 19, 2021 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.