PKG, BIOPSY FORCEPS, SPOON W/PIN, P/N 0250080291. Laparoscopic Manual Instruments are intended f...
FDA Device Recall #Z-0668-2015 — Class II — November 17, 2014
Recall Summary
| Recall Number | Z-0668-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 17, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Stryker Endoscopy |
| Location | San Jose, CA |
| Product Type | Devices |
| Quantity | 180,573 units |
Product Description
PKG, BIOPSY FORCEPS, SPOON W/PIN, P/N 0250080291. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason for Recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lot / Code Information
42337 43487 844401 844761 845042 848259 940167 941850 942654 942773 945445 945980 1040392 1041058 1042090 1042644 1042646 1140811 1141158 1142154 1142667 1143974 1144313 1144620 1145338 1145890 1145973 1242131 1243368 1244190 1244965 1341566 1342767 1342768 1343714 1345270 1345728 1346217 1440423 1440471
Other Recalls from Stryker Endoscopy
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2016 | Class II | PneumoSure XL High Flow Insufflator and Pneumo... | Mar 28, 2016 |
| Z-1831-2015 | Class II | SERFAS 90 degree Energy Probe, Part Number 279-... | Jun 3, 2015 |
| Z-0631-2015 | Class II | PKG, PENNINGTON FORCEPS, P/N 0250080242. Lapa... | Nov 17, 2014 |
| Z-0646-2015 | Class II | PKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250... | Nov 17, 2014 |
| Z-0679-2015 | Class II | PKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparo... | Nov 17, 2014 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.